The AffloVest is intended for use in the electromagnetic environment specified
below. The customer or user of the AffloVest should ensure that it is used in such an
environment.
INTERFERENCE
IEC 60601-TEST
IMMUNITY TESTS
LEVEL
Electrostatic
±6 kV Contact
discharge (ESD) as
Discharge
per IEC 61000-4-2
±8 kV Air Discharge
Fast Electrical
±2 kV for Power Supply
Transients/Bursts as
Lines
per IEC 61000-4-4
±1 kV for Input/Output
Lines
Surge Voltages
±1 kV Normal-Mode
(Surge) as per IEC
Voltage
61000-4-5
±2 kV Common-Mode
Voltage
EVoltage Dips,
<5% Ut (>95% dip in
Short- Term
Ut) for 0.5 cycle
Interruptions and
40% Ut (60% dip in Ut)
Fluctuations in
for 5 cycles
Supply Voltage as
per IEC 61000-4-11
0% Ut (30% dip in Ut for
25 cycles
<5% Ut (>95% dip Ut
for 5 s
Power Frequency
3 A/m
(50 Hz) Magnetic
Field as per IEC
61000-4-8
Note: Ut is the mains AC voltage prior to application of the test level.
24
ELECTROMAGNETIC
COMPLIANCE
ENVIRONMENT
LEVEL
GUIDELINES
±6 kV Contact
Floors should be wood,
Discharge
concrete, or ceramic tile.
If floors are covered with
±8 kV Air Discharge
a synthetic material, the
relative humidity should be
at least 30%
±2 kV for Power Supply
Mains power quality
Lines
should be that of a typical
commercial or hospital
±1 kV for Input/Output
environment
Lines
±1 kV Normal-Mode
Mains power quality
Voltage
should be that of a typical
commercial or hospital
±2 kV Common-Mode
environment
Voltage
<5% Ut (>95% dip in
Mains power quality
Ut) for 0.5 cycle
should be that of a typical
commercial or hospital
40% Ut (60% dip in Ut)
environment. If the user
for 5 cycles
of the AffloVest requires
0% Ut (30% dip in Ut)
continued operation
for 25 cycles
during interruptions in
the power supply, it is
<5% Ut (>95% dip Ut)
recommended that the
for 5 s
AffloVest be powered from
an uninterruptible power
supply or battery
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a location
in a commercial or hospital
environment
The AffloVest is intended for use in the electromagnetic environment specified
below. The customer or user of the AffloVest should ensure that it is used in such an
environment.
INTERFERENCE
IEC 60601-TEST
COMPLIANCE
IMMUNITY
LEVEL
LEVEL
TESTS
3 V rms
Conducted RF
3 V
150 kHz to
rms
interference as per
80 MHz
IEC 61000-4-6
Radiated RF
3 V/m 80MHz to
3 V/m
interference as per
2.5 GHz
IEC 61000-4-3
NOTE 1: The higher frequency range applies at 80 MHz and 800 MHz.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects, and persons.
(a) The field strength from fixed transmitters, such as base stations for radio telephones and
mobile terrestrial radio equipment, amateur radio stations, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic envi-
ronment in terms of fixed RF transmitters, an electromagnetic site survey should be considered. If
the measured field strength in the location in which the AffloVest is used exceeds the applicable
compliance level specified above, the AffloVest should be observed to verify normal operation.
If abnormal performance features are observed, additional measures may be necessary, such as
re-orienting or relocating the AffloVest.
(b) The field strength should be less than 3 V/m over the frequency range from 150 kHz to 80 MHz.
ELECTROMAGNETIC ENVIRONMENT
GUIDELINES
Portable and mobile RF communications
equipment should be used no closer to any
part of the AffloVest, including the cables,
than the recommended distance calculated
from the equation applicable to the transmit
frequency.
Recommended distance:
d = 1.2 m P
d=1.2 m P for 80 MHz to 800 MHz d=2.3 m P
for 800 MHz to 2.5 GHz
Where P is the maximum output power rating
of the transmitter in watts (W) according
to the transmitter manufacturer and d is
the recommended distance in meters (m).
The field strength of fixed RF transmitters,
as determined by an electromagnetic site
survey(a) should be less than the compliance
level in each frequency range(b). Interference
may occur in the vicinity of equipment marked
with the following symbol.
25