Greater Goods GBF-1719-B/0383 Gebruikershandleiding - Pagina 11

Blader online of download pdf Gebruikershandleiding voor {categorie_naam} Greater Goods GBF-1719-B/0383. Greater Goods GBF-1719-B/0383 11 pagina's. Body fat analyzer

Appendix
Table 2
Guidance and manufacturer's declaration – electromagnetic immunity
IMMUNITY test
IEC 60601-1-2 test level
Electrostatic
±8 kV contact
discharge (ESD)
±2 kV, ±4 kV, ±8 kV, ±15 kV air
IEC 61000-4-2
Electrical fast
transient/burst
Not application
IEC 61000-4-4
Surge IEC
Not application
61000-4-5
Voltage dips,
short interruptions
and voltage
Not application
variations on
power supply
input lines
IEC 61000-4-11
Power frequency
magnetic field
IEC 61000-4-8
Conduced RF
Not application
IEC61000-4-6
10 V/m
Radiated RF
80 MHz – 2,7 GHz
IEC61000-4-3
80 % AM at 1 kHz
NOTE
U
is the a.c. mains voltage prior to application of the test level.
T
18
Compliance level
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Not application
Not application
Not application
30 A/m
30 A/m
50Hz/60Hz
50Hz/60Hz
Not application
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Table 3
Guidance and manufacturer's declaration - electromagnetic Immunity
Modulation
Test
Band
Service
Frequency
(MHz)
Radiated RF
(MHz)
IEC61000-4-3
(Test specifications
385
380
TETRA
Pulse
for ENCLOSURE
-390
400
modulation
PORT IMMUNITY
18 Hz
to RF wireless
communications
FM
equipment)
450
380
GMRS 460,
± 5 kHz
-390
FRS 460
deviation
1 kHz sine
710
Pulse
704
LTE Band
modulation
745
-787
13,17
217 Hz
780
GSM 800/
Pulse
800
810
900,TETRA
modulation
-960
800,iDEN
18 Hz
870
820,CDMA
850,LTE
930
Band 5
GSM 1800;
Pulse
1720
1700
CDMA1900;
modulation
-1990
GSM 1900;
217 Hz
1845
DECT;
LTE Band
1, 3,4, 25;
1970
UMTS
Bluetooth,
Pulse
2450
2400
WLAN,
modulation
-2570
802.11
217 Hz
b/g/n,RFID
2450,LTE
Band 7
WLAN
Pulse
5240
5100
802.11 a/n
modulation
-5800
5240
217 Hz
5785
Appendix
Modulation
Distance
IMMUNITY
(W)
(m)
TEST LEVEL
(V/m)
1,8
0.3
27
2
0.3
28
0.2
0.3
9
2
0.3
28
2
0.3
28
2
0.3
28
0.2
0.3
9
19