DayTronic 3770 Gebruiksaanwijzing - Pagina 5
Blader online of download pdf Gebruiksaanwijzing voor {categorie_naam} DayTronic 3770. DayTronic 3770 16 pagina's. Dc strain gage panel meter
Model 3770
2
C
O N N E C T I O N S
J1 – Shunt Resistor Connections. Two terminal pins are provided via the J1 rear connector
for installation of an external shunt resistor. Terminal connections are used to install a user
defined resistor as determined by the sensor's calibration data sheet or pre-determined by
the user. When an externally installed resistor is used, the ANALOG CONTROL switch 5 is
required to be in the External position (Switch 5 DOWN).
The 3770 unit comes with an internally installed 59K Ohm resistor, as standard, and
Note:
enabled when the front panel ANALOG CONTROL Switch# 5 is in the Internal position
(Switch 5 UP).
When a sensor has an integrated shunt resistor, the ANALOG CONTROL switch
Note:
setting needs to be in the External position and a shorting wire in place of the external
shunt resistor position. This provides the electrical path for the Calibration Sense line to
enable the sensor's internal shunt resistor when the + or – Shunt Calibration rear terminals
or push buttons are activated.
NOTE:
Daytronic Shunt Calibration circuit, when activated, will place the Shunt resistor across
the + Sense Line (for positive shunt) or the – Sense Line (for negative shunt) with the
connection of the Sensor's Calibration Sense line located on the J2 connector - which is
recommend to be connected to the + Signal line. When not activated, the Calibration Sense line
is buffered as to not affect the sensor's operational characteristics.
DC STRAIN GAGE PANEL METER
External Shunt Resistor Connections
J1
SHUNT Resistor – User Provided
(Note: ANALOG CONTROL switch
5 must be in the External position)
Page 4