DayTronic IL78A Handmatig - Pagina 13
Blader online of download pdf Handmatig voor {categorie_naam} DayTronic IL78A. DayTronic IL78A 17 pagina's. In-line ac strain gage conditioner ac strain gage conditioner module
3. CALIBRATION
Remote Calibration Check: The instrument can be placed in the calibration mode (positive or negative) by
shorting terminals Power Common and SHP for positive or SHN for negative on the lower I/O connector.
Figure 4 indicates two methods of remotely entering the calibration mode (external switch, or push
button control). The Remote Cal function provides a convenient method of periodically monitoring
calibration of the instrument in the positive or negative realm.
Master/Slave Connections. : When more than one IL78A is being used in a measurement setup
(instruments are closely mounted or the transducer cabling is in a common conduit or raceway),
beat frequencies may be produced from the 3-kHz oscillators used in the instruments to develop the
excitation. To prevent beat frequencies from occur-ring, one unit can be designated the master, and
the remaining units can be driven from the oscillator contained in the master unit. The remaining
units are designated as slave instruments. To perform master/slave wiring, refer to Figure 7.
AC Strain Gage Conditioner Module
(cont.)
Switch 1 - Mode -
Switch 2 - Volts -
Switch 3 - Current -
Switch 4 - Filter -
Switch 5 - Filter -
Switch 6 - Sync -
Switch 7 - Zero Adj. -
Switch 8 - Aux Output -
Coarse Zero -
balance control marked for calibration default position
Fine Zero -
Coarse Span -
contr ol marked for calibration default position
Fine Span -
Symmetry -
Phase -
Over Range -
Power -
Fig. 2 Front Panel Settings and Indicators
current (I) or voltage analog output
+/- 5 or +/- 10 Vdc when mode is voltage
4-12-20 or 4-20mA when mode is current
650 Hz or 100 Hz at 3 db, for 100 Hz switch 5 must be set down
10 Hz or 100 Hz at 3 db, for 100 Hz switch 4 must be set down
excitaton clock to slave from a master module or to be master
extended (100%) or normal (25%) zero authority
SYC terminal output for use by other modules
16 position switch adjustment for stepped zero
18 turn potentiometer for fine zero balance control
16 position switch adjustment for stepped gain
18 turn potentiometer for fine gain - span control
adjust the negative output span to be equal
to the positive output
adjustment for phasing the AC modulated signal to its
highest level at load
indicates when the analog output is 2% greater
than mode selected
indicates the power input voltage is ON
IL78A
Calibration IL78A.3