3) Operation and Installation

3.1) General

The general sensor connector configuration is given below;
Cable Code/Pin Configuration:
• Red
: V +
Power supply voltage +6 to +20 VDC
• Black
: Ground
Power GND
•X-Axis: Yellow : Signal
Analog output voltage signal for single ended
• Y-Axis:
Blue : Signal
Analog output voltage signal for single ended
• Z-Axis : White : Signal
Analog output voltage signal for single ended
WARNING
Never connect the power supply and/or the power ground to yellow, blue, and/or
white cables.
Never connect the power supply to the power ground. Always use a clean power
source and check the voltage range.
4)

Sensor Static Calibration Verification

Using gravity, voltage values are measured in the + and - directions, providing a value of 1 g.
The measurement should be made as follows;
When the acceleration value of 3000SE series sensors is entered into the data
acquisition system, the sensor shows +1 g with the effect of gravity, which is
in the direction of the axis to be calibrated.
When the sensor is positioned in the opposite direction to the axis to be
calibrated, the arrow shows -1g as shown below under the effect of gravity.
Using gravity, the voltage values that provide 1 g in the + and - directions are
measured and compared with the catalog value. The calibration value should
be close to the catalog value with a 10% tolerance. Sensor catalog sensitivity
values are given in Table 1.
7