process interrupted between stage 4 and 5, then the machine will take the previous 4 points
data as the calibrated value, and becomes 4-points calibration.
7. During calibration, the calibration set point could not be changed unless interrupting the
process by pressing the "Start/stop" and "mains" knob as indicated in the step 3.
8. Always using the same type of the block to avoid the discrepancy. The calibrated
displayed value only accurately indicates the value of the block/vial that actually measured
from external temperature measurement device.
For DB-2-200-02:
Figure 4: DB-2-200-02 Overview: Ch1: The left hand side (with six-hole block); Ch2:
The right hand side with 20 hole block)
Figure 5: The front panel view of the DB-2-200-02
Please make sure that the correct channel corresponding to the designated block
before to do the operation. The DB-2-200-02comes with additional LED display
and "sel" button for users to select the channel. All operation and calibration is
the same as in DB-3-200-01, but before to do that, please press the "sel" button to
select the channel (the LED light will be ON when the channel got selected).