Gainlab Audio GLA-TC2 Kalibratiehandleiding - Pagina 11

Blader online of download pdf Kalibratiehandleiding voor {categorie_naam} Gainlab Audio GLA-TC2. Gainlab Audio GLA-TC2 18 pagina's.
Ook voor Gainlab Audio GLA-TC2: Handmatig (13 pagina's)

Gainlab Audio GLA-TC2 Kalibratiehandleiding
5.
STEP - SC LEVELLING
Starting pont:
Input signal : +0dBu 1kHz sinewave, on both channels Threshold 0dB,
mode – HiComp, UNLINK
5.1 step – Stage One
Measurement process: Stage ONE left
Measure: between TP8 (gnd point on volt meter) and TP15
Adjust: VR5
Target: measure +1.1V on a volt meter
Now, turn the VR7 counterclockwise until the end point (appx. 6 turnes)