Gait up PhysiGait Lab Gebruikershandleiding - Pagina 4
Blader online of download pdf Gebruikershandleiding voor {categorie_naam} Gait up PhysiGait Lab. Gait up PhysiGait Lab 18 pagina's. User manual and outcome parameters
3. Gait recording detailed description
3.1
Attachment
The PhysiGait Lab is delivered with 2 rubber clips as default accessory. Elastic foot straps
can be bought from the web-shop (shop.gaitup.com) if needed, for example to measure
barefoot gait.
To insert the Physilog®5 into the rubber clip, slide the sensor into the clip on one side then
carefully pull the rubber of the opposite side to cover the edge of the Physilog®. The USB
entry of the Physilog® should be oriented in front of the clip opening, so that the sensor can
be charged without removing the clip.
For gait measurement the sensor can be fixed in any orientation on the foot, either in the
shoelaces or on the outer side of the shoe (on the inner side there is a risk of disturbing the
gait). The computed gait parameters will not be influenced by the sensor position. Fix the
sensor such that it doesn't move with respect to the foot / shoe.
3.2
Start the recording from the sensors
To start the sensor, push the main button for about 3 seconds. The LED becomes green
and stops again, at this moment you can release the button. It doesn't matter which
Physilog® you turn on first.
For firmware version 1.0.4, the "master" sensor will start to blink green immediately, the
slave sensor shows a continuous green light until it has obtained the synchronisation signal
which starts its recording ("Initializing" in the illustration above).
For later firmware versions (v1.1.0 upwards, released June 2017), both sensors start to
blink from the start, when they are synchronized the blinking is simultaneous. To better see
if they are synchronized it is recommended to start the sensors with a short lag between the
starts.
In all cases wait until the two sensors blink green simultaneously before starting the
movement. The best practice is to start the sensors when they are already attached to the
foot. Otherwise turn them on just before fixing on the subject and ask the subject to stand
still for 5 seconds with the Physilog
PhysiGait Lab user manual v1.3.1 Date: 29.04.2019
Copyright ©2018, All Rights Reserved - Gait Up SA. CH-1015 SWITZERLAND
EPFL Innovation Park, Bat C. CH-1015 Lausanne
®
on the feet to allow proper auto-calibration of the system.
Gait Up SA
www.gaitup.com
TVA CHE-145.136.760
4