The AUTOSCALER® should NOT be used on or near anyone fitted
with a CARDIAC PACEMAKER.
Do not remove the cover from the unit. There are NO user
serviceable parts inside. Allow only qualified technicians to service
your AUTOSCALER®.
DO NOT AUTOCLAVE the detachable HANDPIECE assembly. A
Federally approved non-immersion WATER BASED cold
sterilization solution may be used on the detachable handpiece
assembly. However, DO NOT get sterilization solutions on the
electrical contacts of the Handpiece CONNECTOR as this will cause
oxidation, and a poor electrical connection. DRY ALL SURFACES
with clean air or clean paper towels.
If your unit is left ON, and the footswitch is accidentally depressed
while unattended, this will permanently damage your unit. Be sure
nothing is touching the operatory air foot switch when not in use.

INFECTION CONTROL

Manufacturer recommends that all operators of this ultrasonic
unit wear at least one pair of surgical rubber gloves, and for extra
prevention wear double pairs of gloves to isolate the operator
from infection, and cross contamination of other patients.
It is recommended that complete eye and face protection be used
during the use of this unit to prevent saliva spray from being
inhaled, and from entering the operator's eyes.
Operator should be careful not to puncture the skin with any
insert tip containing saliva from patients. Insert tips should be
sterilized after each patient according to the insert manufacturer's
instructions.
DO NOT AUTOCLAVE the detachable HANDPIECE assembly. A
Federally approved non-immersion WATER BASED cold
sterilization solution may be used on the detachable handpiece
assembly. However, DO NOT get sterilization solutions on the
electrical contacts of the Handpiece CONNECTOR as this will cause
oxidation, and a poor electrical connection. DRY ALL SURFACES
with clean air or clean paper towels.
Manufacturer recommends the use of a disposable handpiece
sleeve/barrier that keeps contamination from getting on the
handpiece assembly. Dispose of sleeve/barrier after each patient.
Ask your local dealer for HIGH SPEED LONG sleeve/barrier.
5

INTRODUCTION

Your South East Instruments AUTOSCALER® Ultrasonic Scaler is a
precision engineered instrument, constructed with the finest of
materials, and requiring no special maintenance. By taking the
following precautions your unit will give you many years of service.
1. Do not place unit on or near any source of heat.
2. Clean the outer surfaces of the AUTOSCALER® with a Federally
approved non-immersion WATER BASED cold sterilization solution.
Dry immediately with clean air or clean paper towels.
3. Unit should be located where a normal amount of air will circulate
freely around the case when in use.
4. The handpiece and Quick Connector are very strong but care
should be taken NOT TO PULL with excessive force.
Basically, the unit consists of three major components: the electronic
generator, the water system, and the handpiece assembly.

ELECTRONIC GENERATOR AND WATER SYSTEM

The electronic generator produces the frequency and the power to
activate a wide variety of interchangeable Cavitron® insert tips. The
water system provides precise control over the water lavage flow
which serves to wash debris from the work area, and to cool the
laminations of the insert tip. The following controls are found on the
front panel of your AUTOSCALER®:
Toggle Switch : To turn unit On and Off.
Indicator L.E.D. Lamp (above the switch) which lights when the
Toggle Switch is set to the ON-I position, indicating the unit is
ready for operation.
Power Control Knob: The Linear Power Control knob adjusts the
output power to the insert tip. The blue
shaded area is for use when using
Cavitron® SLIMLINE™, FSI™ or modified
Periodontal inserts. It is recommended
the Power not be set higher than 12:00
even if no mist appears at the tip.
Slimline inserts do not easily mist
the water, and are very fragile at higher
power settings. FOLLOW INSERT
MANUFACTURER'S COMPLETE
INSTRUCTIONS.
6