Afinion AS100 Beknopte handleiding - Pagina 4
Blader online of download pdf Beknopte handleiding voor {categorie_naam} Afinion AS100. Afinion AS100 4 pagina's.
Afinion
ACR
TM
Control Testing
Read the entire Afinion™ ACR Control Package Insert before use.
How often do I have to run controls?
•
With each new shipment of ACR kits.
•
With each new lot of ACR kits.
•
At least every 30 days.
•
When training new users.
•
Anytime an unexpected test result is obtained.
How should I use the Afinion™ ACR controls?
•
The controls can be used directly from the refrigerator without equilibration to room temperature.
•
Mix the control well by inverting the vial 8-10 times before collecting a sample.
•
Analyze the control using the procedures described on page 2 (Specimen Collection) and
page 3 (Running Samples on the Analyzer).
•
Compare the test results with the values listed on the front page of the Afinion™ ACR Control
Package Insert.
What do I do if Afinion™ ACR Control results are not within the
acceptable range?
•
Do not analyze any patient samples.
•
Check the control vial label to make sure it is not expired.
•
Ensure that the control has not been used for more than 8 weeks.
•
Verify that the controls and test cartridges have been stored correctly.
•
Verify that there is no visual sign of bacterial or fungal growth in the control vial.
Correct any procedural error. Re-test the control.
If the control values are still not within acceptable range, repeat the test using a new vial of control.
If the control results are still not acceptable, call Technical Support.
Technical Support?
Call 1-866-216-9505. This is a toll free number.
Available for use only in the United States of America.
Axis-Shield PoC AS
an Alere company
P.O. Box 6863 Rodelokka
NO-0504 Oslo, Norway
alere.com
afinion.net
© 2012 Alere. All rights reserved. The Alere Logo, Alere and Afinion are trademarks of the Alere group of companies. PN: 1115297 Rev. A 2012/10
Quick Guide
4