Sample Cup
Stainless steel cups are
recommended. Plastic cups
absorb moisture over time.
Weighing Pan
Air Cylinder
Base Block Bore
Figure 4: Sample cup and lower block assembly
• Figure 4: assembly illustrates the proper positioning of the
parts in the lower block. The air cylinder is placed into
the bore of the base block. It should slide into the block
easily and settle at the bottom of the bore. The weigh
pan connects to the scale and holds the sample cup. For
best results, verify cleanliness of all parts.
2. AquaLab VSA Performance Verification overview—
• AquaLab VSA performance and cleanliness should be
checked frequently; before each isotherm test is
recommended.
• Press until you are at the Measurement Tab.
• Use AquaLab Verification Standards to verify performance
or adjust linear offset.
V S A Q u i c k S t a r t G u i d e
Sample Cup Lid
Only used for sample storage and
must be removed during tests.