Decagon AQUA LAB VSA Snelstarthandleiding - Pagina 5

Blader online of download pdf Snelstarthandleiding voor {categorie_naam} Decagon AQUA LAB VSA. Decagon AQUA LAB VSA 12 pagina's. Vapor sorption analyzer

Sample Cup
Stainless steel cups are
recommended. Plastic cups
absorb moisture over time.
Weighing Pan
Air Cylinder
Base Block Bore
Figure 4: Sample cup and lower block assembly
   •  Figure 4: assembly illustrates the proper positioning of the 
       parts in the lower block. The air cylinder is placed into 
       the bore of the base block. It should slide into the block 
       easily and settle at the bottom of the bore. The weigh 
       pan connects to the scale and holds the sample cup. For 
       best results, verify cleanliness of all parts.
2. AquaLab VSA Performance Verification overview—
  •  AquaLab VSA performance and cleanliness should be
    checked frequently; before each isotherm test is 
    recommended.
  •  Press   until you are at the Measurement Tab.
  •  Use AquaLab Verification Standards to verify performance
    or adjust linear offset.
V S A Q u i c k S t a r t G u i d e
Sample Cup Lid
Only used for sample storage and 
must be removed during tests.