Major Science MP-300 Gebruiksaanwijzing - Pagina 14
Blader online of download pdf Gebruiksaanwijzing voor {categorie_naam} Major Science MP-300. Major Science MP-300 16 pagina's. Powerpoint power supply
Error Messages
Err1; power
supply stops;
alarm sounds.
Err2; power
supply stops;
alarm sounds.
High voltage
application is set to
run on a very low
current
Current exceeds
setting
Voltage exceeds
setting
DISABLE No Load alarm on the
Display Screen
-
Check that the correct
running buffer and buffer
concentration have been
used
-
Check that cables are
connected properly
-
Switch power on / off and
then restart application
-
If the power supply cannot be
restarted, disconnect the
power supply from the mains
outlet and contact Technical
Service
-
Check that the correct
running buffer and buffer
concentration have been
used
-
Check that cables are
connected properly
-
Switch power on / off and
then restart application
-
If the power supply cannot be
restarted, disconnect the
power supply from the mains
outlet and contact Technical
Service
14