7.1.2. Insulation measuring procedure:
1. Connect to test cable to test object.
2. Turn the function rotary switch to the test voltage from
3. When pressing and holding the backlight button then the
4. Be sure that the cable connecting the test object does not
If pressing the TEST/ STOP button, the LCD displays the
warning picture show in below
Note: The test cannot be performed. To test remove the
exterior voltage.
5. Press the TEST / STOP button for testing continuously.
6. Press the TEST / STOP button for discharging.
7. Read the test value from LCD display.
10 10
1000V, 2500V, 5000V or 10000V respectively.
display will shows the backlight function.
have a voltage (above AC30V or DC30V). The LCD will
display the voltage picture as show in the AC/ DC voltage.
While test is running, beeper activates to remind -9-
operator that test is underway. After the set test time is
completed (see Function 2: the test time setting), the test
stops and system will automatically hold the test value. To
read the test value on the LCD display.