Instructions. Ambient model
D. Programming, 100 are available from 0 – 99
Set all parameters as previous details.
After setting all your required parameters.
Press
Function
and select memory to your desired number by pressing
Once done whilst pressing
To recall a program Press
Once done whilst pressing
NOTE. The last run data (program must be run) will be kept, after turning off the Centrifuge power

Simple problems

Speed will not go beyond 2000 Rpm.
.
The Electronics have set the Speed to safe mode
.
Check the rotor bottom has magnets fitted
.
Check correct rotor is in use against program chosen
.
If not then
Noisy
.
Unbalanced tubes, see Rotor pages
.
Lid seal needs replacing
.
Rotor Lid not fitted
Lid support fail
.
Spring has failed.
A
Rotor,
Store
Function
Function
Press
Function
and memory to desired program number
Function
Press
Contact our Service Department
Contact our Service Department
B
. Speed or Rcf(G), Accel, Decell.
Store.
Recall.
C
Time
9 9
Memory
Memory
Memory
Recall
Memory
buttons
14