CORNING LSE 6781-4 Gebruiksaanwijzing - Pagina 3
Blader online of download pdf Gebruiksaanwijzing voor {categorie_naam} CORNING LSE 6781-4. CORNING LSE 6781-4 10 pagina's. Digital microplate and orbital shakers
1.0 Overview
The Corning® LSE™ Digital Microplate and Orbital Shakers are intended for shaking microplates, tubes, bottles, flasks, dishes, and
other laboratory vessels. The shakers are driven by asynchronous motors, which enable silent operation and constant shaking speed
independent of the load or power supply fluctuations.
Each shaker device consists of two main components:
Motor with eccentric mechanism
Control electronics
The motor drives the shaker's eccentric mechanics and generates shaking effect. The electronics control the motor RPM, TIME, and
keyboard functions.
2.0 Safety Precautions
Do not use the shaker near water. Ensure that water does not get on the device, particularly during cleaning.
If the unit is used in a manner not specified in this manual, then the unit may become unsafe to operate and may void
the warranty.
This product must be connected to a grounded power outlet for safe functioning.
Use only the power cord supplied with the product.
The equipment will not restart if the main power source is interrupted.
The equipment will restart in cases of mechanical interruption.
Do not expose the shaker to aggressive vapors (e.g., organic solvents).
Do not insert any objects into the shaker (such as a screwdriver or metal rods).
Never operate the shaker without the platform securely attached.
Hazardous samples should be shaken only in appropriate containment vessels.
Do not shake flammable or explosive samples.
The electrical warning symbol indicates the presence of a potential hazard which could result in electrical shock.
Do not put your fingers between the shaking platform and the housing. Danger of mechanical injury.
3 .0 Technical Specifications
The shaker housing is made from a high-grade cold rolled steel plate and painted with a highly resistant polyurethane lacquer.
Microplate Shaker (Cat. Nos. 6780-4, 6781-4, 6782-4)
Electrical Requirements
120V
230V
Motor Power
Fuse
120V
230V
RPM Regulation
Shaker Orbit
Timer
Maximum Load
Dimensions (L x W x D)
Weight
Operating Conditions
120V ± 10%, 50/60Hz
230V ± 10%, 50Hz
15W
2 x T 1A L 250V
2 x T 0.25A L 250V
Digital, load independent, from 100 to 1,200 (1,400, 120V version) RPM in 10 RPM steps
3 mm
30 sec. to 99 min. 50 sec. in 10-second steps (under 10 min. in 1-second steps), timer HOLD function
0.6 lbs. (0.3 kg)
6.6 x 10.98 x 6 in. (16.8 x 27.9 x 15.3 cm)
9.5 lbs. (4.3 kg)
5°C to 40°C, 85% RH, non-condensing
Instruction Manual
|
1