a&p instruments UX 3081 Gebruiksaanwijzing - Pagina 13

Blader online of download pdf Gebruiksaanwijzing voor {categorie_naam} a&p instruments UX 3081. a&p instruments UX 3081 20 pagina's. Moisture analysers

Operating instructions for moisture analysers
UX 3081
UX 3081WQ
Moisture analysis process
The device should be connected to the mains supply and switched on with a drying tray. A weight value is
displayed.
Set scales to zero using the 'tare' button,
display '0.000 g'.
1.
2. Swing the heater over the sample, switch on the
heater switch and the drying phase begins.
Header data is now printed for devices with a built-in
printer. The sample is dried until it has reached a
constant weight; the heater then switches off
automatically. The time and measurement value is
printed for devices with a printer.
If you do not wish to make any measurements,
please always ensure that the heater is swung back
to its rearmost position!
3. Measurement values are displayed for the entire length of timethe heater is positioned above the sample.
!
The device should not be moved from its position during measurement, as this causes measurement errors.
The external printer prints the following
at the start of the measurement:
Model, device ID,
date, time,
heater temperature,
timer or shut-off parameters, initial weight.
Please ensure drying trays have cooled down before using for measurement.
If the measurement ends before
it is complete,
the display shows:
For substances which are difficult to remove from the
tray after being dried, we recommend using foil trays,
which can be made quickly and economically from
aluminum foil using the ULTRA X foil press.
a&p instruments, Kluckhofer Weg 61, D-32657 Lemgo, Tel.:05232 97780, Fax.: 977820
E-Mail: [email protected], Internet: www.apinstruments.de
Evenly distribute approx. 50 - 200 g
sample matter in the tray and wait
for the readout to stop.
And after the device has finished measuring, once the
heater has been swung back to its rearmost position:
time
mode
value (weight)
unit
green
Measurement value,
measurement time,
substance,
initials
esc
mode
value (weight)
to sample weight.
glow if
red
heater is on
flashes if
measurement
ends
LED´s are off
if switch „off"
time
unit
Seite 13