TROUBLE SHOOTING
If you experience difficulties in operating your Kardia products,
refer to the troubleshooting guide below or contact technical
support at [email protected].
Problem:
Solution
My Kardia Mobile is
Option 1: Ensure that the Kardia phone app
not working.
has access to the phone's microphone. On
the iPhone, go to Settings and tap the Kardia
app. Tap the microphone toggle.
Option 2: Change the battery. Remove Kardia
Mobile from the attachment plate (if applicable),
and remove the battery door with a pen, pencil,
or other similarly shaped object into the cutout
next to the battery door to pop the battery off.
Problem:
Solution
I have a lot of artifact,
Option 1: Ensure that the "Enhanced Filter"
noise, or interference in
is on.
my recording.
Option 2: Ensure that your arms and hands
remain still during recordings.
Option 3: Clean the electrodes on the Kardia
Mobile with an alcohol-based sanitizer.
Option 4: If your hands are very dry, use a
water-based lotion before recording.
Option 5: When recording, relax your arms
and hands to reduce muscle noise. Rest the
forearms and hands on a flat surface.
Option 6: If you wear hearing aids, turn them
off prior to recording.
Problem:
Solution
The ECG rhythms appear
Option: Ensure that the left hand contacts the
upside down.
electrode closer to the top of the smartphone
and the right hand contacts the electrode to
the bottom of the smartphone.
CAUTIONS :
General:
DO NOT store in extremely hot, cold,
humid, wet, or bright conditions.
DO NOT expose to strong electro-
magnetic fields.
DO NOT take recordings in close
vicinity to other equipment emitting
ultrasonic acoustics.
DO keep components out of reach
of children.
DO use this device to record heart
rate and heart rhythm only.
DO NOT use the electrode on a
portion of the body with too much
body fat, body hair or very dry skin,
a successful recording may not be
possible.
AliveCor makes no warranty for any
data or information that is collected
erroneously by the device, or misuse
or malfunction as a result of abuse,
accidents, alteration, misuse, neglect,
or failure to maintain the products
as instructed. Interpretations made
by this device are potential findings,
not a complete diagnosis of cardiac
conditions. All interpretations should
be reviewed by a medical professional
for clinical decision-making.
Kardia Mobile
DO NOT use with a cardiac pace-
maker, ICDs, or other implanted
electronic devices.
DO NOT continue use until further
instructed by a physician if your
skin is irritated or inflamed around
the sensor.
DO NOT drop or bump with
excessive force.
DO NOT use to diagnose heart-
related conditions.
DO NOT wear during magnetic
resonance imaging (MRI), cautery
and external defibrillation procedures.
After ECG analysis, the app may
incorrectly identify ventricular flutter,
ventricular bigeminy, and ventricular
trigeminy heart conditions as
unreadable. Please consult with
your physician.
CAUTION: AliveCor does not
guarantee that you are not expe-
riencing an arrhythmia or other
health conditions when labeling an
ECG as normal. You should notify
your physician for possible changes
in your health.
KARDIA MOBILE
SPECIFICATIONS
Battery: Coin Cell
Storage Conditions: Original
package under normal room
temperature and humidity
ELECTROMAGNETIC & OTHER
INTERFERENCES
Kardia Mobile has been tested and
deemed in conformance with the
relevant requirements in EN60601-1
-2:2007 Class B for Electromagnetic
Compatibility (EMC).
EQUIPMENT SYMBOLS
Type CF Applied Part
Do not dispose with
household waste
Read instructions
before use
Manufacturer
Temperature range
Humidity range
Model number
Serial number
European Authorized
Representative
ADDITIONAL INFORMATION
For more detailed troubleshooting
and technical information, please
visit: https://www.alivecor.com/ /