A-M Systems 1700 Handmatig - Pagina 8
Blader online of download pdf Handmatig voor {categorie_naam} A-M Systems 1700. A-M Systems 1700 16 pagina's. Four-channel differential ac amplifier
General Notes
Any amplifier channel not currently in use should have its
position to protect the amplifier inputs. This grounds the inputs to the differential
amplifier circuitry.
Set the
,
GAIN
LOW CUT-OFF
frequency content of the signal to be recorded.
While recording from biological preparations take care to keep all instrument cables as
far away as possible from the recording situation. This will assist in maintaining proper
grounding and shielding to insure a minimum of electrical interference.
Input cables are available to connect the amplifier to extracellular electrodes and/or
stimulators. These cables will attach to either the
Additional cables can be ordered (catalog #692000, #701700). One end of each cable
is left open to allow for maximum flexibility. The pin assignments for the connectors
and the cables are as follows:
Pin
A
B
H
D
E
NOTE: Pin H can be configured to be as a ground, rather than a driven shield. A
driven shield is used with the
the cable, thus increasing common mode rejection. The shield is driven by a low
impedance source with a differential signal voltage from the amplifier. This driven
shield (the silver braid) should not be connected to ground, as this will cause
noise in the input signal. Instructions on how to convert pin H to a ground or shield
are detailed on the next page. Each channel can be configured independently.
The
connector shield is connected to the system ground internally. Therefore,
STIMULUS
any ground-referenced or isolated stimulator can be used with the
A-M Systems
Operating Instructions
filter,
HIGH CUT-OFF
Wire
Black
White
Shield
connector to minimize the effect of capacitance on
INPUT
131 Business Park Loop, P.O. Box 850 Carlsborg, WA 98324
Telephone: 800-426-1306 * 360-683-8300 * FAX: 360-683-3525
E-mail: [email protected] * Website:
switch in the
MODE
filter, and
filter according to the
NOTCH
or the
INPUT
STIMULUS
INPUT
Signal +
Signal -
Driven Shield*
not used
not used
STIMULUS
http://www.a-msystems.com
STIM
connector.
STIMULUS
+
-
Ground
not used
not used
connector.
4