A-M Systems 3000 Gebruiksaanwijzing - Pagina 4

Blader online of download pdf Gebruiksaanwijzing voor {categorie_naam} A-M Systems 3000. A-M Systems 3000 18 pagina's. Ac/dc differential amplifier

A-M Systems 3000 Gebruiksaanwijzing

Instrument Features

The AC/DC Differential Amplifier, Model 3000 is a single-channel, differential amplifier
that can be used with or without a head stage probe. The instrument is designed for
low-noise recordings from excitable tissue. It is intended for extracellular recording
and/or stimulating in conjunction with microelectrodes. It can be used in a number of
research or teaching applications requiring extracellular neurophysiological recording
from excitable tissue, such as nerve, muscle (EMG), EEG, EKG, and ERG recordings.
The instrument is not intended for clinical measurements using humans.
The Model 3000 contains a high-gain, low-noise differential amplifier stage followed by
low frequency, high-frequency, and notch filters. Three operating modes are available
to accommodate recording, stimulating, and verification of electrode impedance.
Record Mode offers six levels of signal gain (x50, x100, x500, x1000, x5000, and
x10,000). Stimulus Mode allows the current passing through the electrode to be
measured during stimulation at six levels of gain (5V/mA, 10V/mA, 0.05V/µA, 0.1V/
µA, 0.5V/µA, 1V/µA). Impedance Mode utilizes an internally calibrated current source
to allow in situ verification of electrode impedance and adjustment of capacitance
compensation. Units that are sold with a head stage are pre-configured to work with
the headstage, and the front panel gain values include the gain of the headstage.

Controls and Connectors

Input Controls
This 15-pin connector receives a signal either directly from the electrodes (non
PROBE:
head stage version) or from the Head Stage probe for further processing.
INPUT SELECT (DIFF-MONO-GND)
input for differential operation, monopolar operation, or grounds the inputs to the
amplifier.
INPUT LEADS (standard):
leads. The black wire is the "+" electrode lead. The white wire is the "-" reference
lead. The silver wire is a driven shield, and should not be grounded! This wire
serves to minimize signal loss due to capacitance.
HEADSTAGE PROBE (optional):
to electrodes.
PROBE+
recording microelectrode; It is the larger 2mm jack).
indifferent lead. Either the
A-M Systems

General Description

(non headstage version only): This switch sets the
Supplied with the Model 3000 is a DB-15 connector with three
The Headstage Probe has three sockets for connection
is used to connect to the electrode (usually a high impedance
connector, or the front panel
PROBE GND
131 Business Park Loop, P.O. Box 850 Carlsborg, WA 98324
Telephone: 800-426-1306 * 360-683-8300 * FAX: 360-683-3525
E-mail: [email protected] * Website:
is used to connect to the
PROBE-
http://www.a-msystems.com
must be
GND
1