A.H. Systems SAS-563B Bedieningshandleiding - Pagina 10

Blader online of download pdf Bedieningshandleiding voor {categorie_naam} A.H. Systems SAS-563B. A.H. Systems SAS-563B 13 pagina's. Active loop antenna
Ook voor A.H. Systems SAS-563B: Bedieningshandleiding (14 pagina's)

A.H. Systems SAS-563B Bedieningshandleiding
A.H. Systems Model SAS-563B Active Loop Antenna

CALCULATIONS

EMISSIONS TESTING
Individual calibration data for the active loop antenna is supplied to comply
with various emissions test requirements. For emissions measurements, add
the correction factor plus cable loss to receiver reading in dBV to convert to
field strength in dBA/meter.
Field Strength(dBuA/m) = SA(dBuV) + MAF(dB/m) + cable loss (dB)
Fictitious electric field strength can be calculated by the following:
© A.H. Systems inc. – Mar 2021
REV E
SA = Spectrum Analyzer or Receiver voltage reading
MAF = Magnetic Antenna Correction Factor
CL = Cable Loss in dB
dBV/m = dBA/m + 51.5
10