Aqua Free Germlyser D Gebruiksaanwijzing
Blader online of download pdf Gebruiksaanwijzing voor {categorie_naam} Aqua Free Germlyser D. Aqua Free Germlyser D 2 pagina's.
7 Maintenance and cleaning
D requires no maintenance within the specified operating time. The outside of the filter
®
The Germlyser
can be cleaned using common surface disinfectants.
D, a small amount of unfiltered water will flow out. Over time, this will lead
When replacing the Germlyser
®
to deposits on and in the adapters. Aqua free recommends replacement of the adapters after a period of
use every 6 to 12 months.
In the case of a deviation from the specifications or a damage of the package, the product should be
returned to Aqua free, along with the Defect Report form delivered with the product (form downloads at the
service page under www.aqua-free.com).
Please note: Aqua free operates its own replacement service, which can be used by facilities with 25 or
more filterpoints of use.
8 Storage
filter must be properly stored in a dry and hygienic environment at room temperature.
The Germlyser
®
9 Returning
reusable filters must be sent back to Aqua free within the agreed time for reprocessing
The Germlyser
®
(form downloads at the service page under www.aqua-free.com).
Aqua free Solutions Ltd.
The Media Centre
7 Northumberland Street
Huddersfield, HD1 1RL, Great Britain
E-Mail:
+44 (0)1484 483045
Phone:
Headquarters:
Aqua free GmbH
Winsbergring 31
22525 Hamburg
Germany
+49 (0)40-46 89 99 90
Phone:
+49 (0)40-46 89 99 99
Fax:
www.aqua-free.com
Instructions for use
Germlyser
®
D
Table of contents
1 Product
2 Certificates / Declaration of conformity
3 Labelling
4 Preparation of commissioning and use of product
5 Disassembly of used filter / filter change
6 Accessories
7 Maintenance / cleaning
8 Storage
9 Return
1 Product
1.1 Product description
The Germlyser® D is a
-marked, point-of-use sterile shower filter for use in high-risk areas in medical
facilities to prevent bacterial nosocomial infectious diseases caused by Pseudomonas aeruginosa,
Legionella pneumophila, Stenotrophomonas maltophilia or Staphylococcus aureus among
immunosuppressed and immunocompromised patients. To achieve this intended use, a specific
combination of retention of these bacteria in the filter membrane structure and specific measurements to
prevent retrogarde contamination with these bacteria on the filter outlet, are necessary.
1.1.1 Designated purpose
The hollow fibre membrane integrated in the Germlyser
waterborne bacteria Pseudomonas aeruginosa or legionella. Users can come into contact with
waterborne bacteria during showering or when cleaning a shower unit and this leaves them open to the
D sterile filter reliably prevents contact with these pathogens and
threat of infection. Using the Germlyser
®
therefore protect against infetion.
1.2 Technical data
Size:
L 206 mm x Ø 43 mm
Retention*:
7 log steps Brev. dim.
Pore size:
0.2 µm
Operating time**:
up to 2 months
Chlorine resistance***: ≤10 ppm
Certification:
* Measured in accordance with the international ASTM F838-15a standard - Standard Test Method for
Determining Bacterial Retention of Membrane Filters Utilized for Liquid Filtration
** Depending on water quality
*** Continuous dosage of ≤ 10 ppm over operating time; short-term (1h) high dosage (400,000 ppm)
for chemical disinfection.
**** The Germlyser
®
D is not suitable for connection to low-pressure systems (e.g. boilers).
D sterile filter guarantees retention of all
®
Minimum flow rate:
12 l/min. at 5 bar, at 25°C
Max. operating pressure****:
5 bar at 60°C
Max. operating temperature:
60°C (at 70°C < 10 min.
over operating time)
Connection:
Quick-release coupling
www.aqua-free.com