Allied Healthcare EPV200 Gebruikershandleiding - Pagina 11
Blader online of download pdf Gebruikershandleiding voor {categorie_naam} Allied Healthcare EPV200. Allied Healthcare EPV200 20 pagina's. Portable battery powered ventilator with oxygen source required
Following is a list of tested cleaning solutions:
Isopropyl Alcohol:
Alconox:
Cetylcide:
Bleach:
Dispose of single patient use items per local biohazard standards.
11
Maintenance and Schedule
The EPV200 should be checked annually to insure proper function. Batteries should be
replaced if they are beyond the expiration date on the battery. Replace with D size
alkaline batteries.
Year 1:
Perform Yearly Verification of function and Calibration procedure. Check battery
expiration date; replace batteries as necessary.
Year 2:
Perform Yearly Verification of function and Calibration procedure. Check battery
expiration date; replace batteries as necessary. Replace disposable O
foot oxygen hose (see Accessories on page 14).
Replace main valve with piston, o-ring, and spring for units manufactured before April
2009. The manufacturing date is coded into the serial number for each unit. The first
four digits identify the year of manufacture, and the next two digits identify the month of
manufacture. Example: the serial number 20090403001 was manufactured in the year
2009 and in the month of April. All repairs should be performed by Allied Healthcare
Products trained and certified technicians.
Year 3:
Perform Yearly Verification of function and Calibration procedure. Check battery
expiration date; replace batteries as necessary.
Year 4:
Perform Yearly Verification of function and Calibration procedure. Check battery
expiration date; replace batteries as necessary.
Year 5:
Perform Yearly Verification of function and Calibration procedure. Check battery
expiration date; replace batteries as necessary.
The unit has a life expectancy of 5 years based on normal usage and proper
maintenance. A comprehensive rebuild will be necessary at this time, and should be
performed by Allied Healthcare Products trained and certified technicians.
70% IPA
1 tablespoon Alconox to 1 gallon H
2 tablespoon Cetylcide to 1 gallon H
10% bleach in H
O
2
10
O
2
O
2
inlet hose with 6
2