1.2.2.3
The pressure array is intended to be used to generate a sequence of pressures in a
specific order. The pressures will be generated using with PS or PSH. This is
determined by the Generation setting. To set up an array, press [A].
min press: MINIMUM PRESSURE is the first pressure that is to be generated.
It may also be the last pressure to be generated if return to start is set to YES.
No limit checking is provided on the minimum or the maximum pressure so care
must be taken to ensure that these values are in the normal operating range of
the PPC.
max press: MAXIMUM PRESSURE is the highest pressure in the sequence
that the PPC is to generate.
# of inc: NUMBER OF INCREMENTS is the number of pressure steps that the
PPC will generate between the minimum and the maximum pressure.
ret to s: RETURN TO START determines if the PPC will return to the minimum
pressure after reaching the maximum pressure. If this is set to NO, then the
pressure array will stop executing after the maximum pressure has been
reached.
reached, the pressure array will continue to generate the same pressures but in
the opposite direction until the minimum pressure has again been reached.
dwell: DWELL is the delay in seconds that the PPC will remain at each pressure
increment after a ready condition has occurred.
received, then this delay will be executed before the next pressure will
be generated. If a delay of 0 (zero) is entered, then the next pressure generation
will not begin until the [C] key has been pressed. This allows you to stay at a
pressure for as long as you like and continue when you are ready.
Once a pressure array has been set up, it can be executed by pressing [F].
1.2.3

REMOTE COMMAND

The remote command can be used to send any command to the PPC. When using the
remote command < -- REMOTE COMMAND -- > will be displayed in the current
pressure window.
commands available. Pressing [ESC] will return you to the pressure monitoring mode.
©2000 DH Instruments, Inc.

PRESSURE ARRAY

If this is set to YES, then after the maximum pressure has been
See the PPC1 Operating Manual for a list of the remote
Page 4
PPC Communication Software User's Manual
After a ready indicator is