AW-Lake EDG-2C-12S Installatie, bediening en onderhoudshandleiding - Pagina 9
Blader online of download pdf Installatie, bediening en onderhoudshandleiding voor {categorie_naam} AW-Lake EDG-2C-12S. AW-Lake EDG-2C-12S 18 pagina's. The edge flow sensor with bluetooth
Output Calibration
Warning modifying these outputs could result in inaccurate analog signals.
This menu is provided to adjust the EDGE analog output to correspond correctly with the
analog output of the input device.
* This menu is intended for use with slight output differences between the EDGE output and customer
input device. These values are factory calibrated before shipping to match meters that are calibrated
yearly.
Select Calibration Type: From this menu both mA and Voltage
outputs may be calibrated. (Figure 6)
Channel A – Voltage output
Channel B – mA output
1. Select which output type to be calibrated; mA or Voltage
2. Press Set Output Low, then enter in the value on user's
device into Record Reading
a. Ex. If EDGE is being calibrated to a multimeter,
enter in reading from multimeter to Record
Reading
3. Press Record button
4. Now press Set Output High to calibrate the high end of the output
5. Enter in High reading from input device to Record Reading
6. Press Record button
7. Verify both Calibration high and low values are correct and press Set Calibration
8. Use back button on phone to navigate to the main menu
*Warning do not do 2 calibrations in a row. After calibration is finished, exit the app and restart
the device.
Figure 6:
Output Calibration
9 |
P a g e