INSTRUCTIONS FOR DPA 02
(2) The reservoir. See Fig 2
The following parts of the reservoir are identified;
A. The pressure relief valve with outlet pressure set at 7.5cm water.
B. The air entry one way valve with arrows indicating direction of air flow.
C. The oxygen supplementation port (metallic nozzle).
D. The 2 litre reservoir bag.
E. Vaporiser.
F. A solid block with four openings and a metallic nozzle.
F
Assembly of reservoir:
1. While standing in front of the vaporiser and facing the DPA 02 anaesthetic machine, hold the
reservoir block so that the metallic nozzle (C) is on the left and pointing away from you.
2. Attach the reservoir block to the input port on the left side of the vaporiser
3. With the reservoir block firmly in place attach as follows:
4. To the rear port attach the air entry valve with the arrows pointing forward
5. To the top port attach the pressure relief valve so that it stands vertically.
6. To the bottom port attach the green reservoir bag
TM
PORTABLE ANAESTHETIC SYSTEM
The Diamedica Portable Anaesthetic
machine DPA 02
components; vaporiser, reservoir, and
Fig. 1
breathing system. It can be rapidly
assembled ready for use as follows:
(1) The vaporiser
Remove the vaporiser from the
container and place it on the wire grill.
Fix the vaporiser to the stand using the
captive screw at the back of the upright
section.
Fig. 2
© Copyright Diamedica (UK) Ltd 2017
TM
has three principal
Page 4 of 13