5-12. Meter Board (PC22) DIP Switch Settings
Factory default
DIP switch set-
tings to display
wire speed as
inches per
minute
DIP Switch Settings:
Display Hold OFF − Set switch 1 so the depressed section is toward OPEN as labeled on the switch.
Displays will not hold values. The voltage displayed while welding is an average reading over a time span of 6 to 8 sec-
onds. If the weld time is less than 8 seconds, the voltage displayed may not be accurate.
Display Hold ON (Factory Default) − Set switch 1 so the depressed section is toward 1 as labeled on the switch.
Displays will hold their last value for five seconds after the trigger is released. The voltage displayed while welding is
an average reading over a time span of 6 to 8 seconds. If the weld time is less than 8 seconds, the voltage displayed
may not be accurate.
OM-278792 Page 20
Factory default
DIP switch set-
tings to display
wire speed as
meters per
minute
2
1
Accessing DIP Switch
1
Lens
DIP switch is located under the lens
next to the displays.
2
Lens Screws
Remove three screws.
Loosen upper right hand screw to
rotate lens enough to access DIP
switch.
Replace three screws removed
earlier.
Tighten upper right hand screw.
.
The two upper screws must
engage the backing plate to
properly secure the front panel.
Ref. 278794-A