BIOMET Clotalyst Handmatig - Pagina 4

Blader online of download pdf Handmatig voor {categorie_naam} BIOMET Clotalyst. BIOMET Clotalyst 12 pagina's. Autologous serum collection system

Preparation of the Clotalys t
®
Autologous Serum
Step 1: Draw
Draw 12ml of citrated blood in a 12ml syringe (1ml
of anticoagulant (ACD-A) and 11ml of blood).
Gently mix the whole blood and anticoagulant upon
collection to prevent coagulation.
Step 2: Load
Remove Clotalyst
reagent syringe from sterile pouch
®
and attach plunger.
Note: Leave grey stopper on opposite end of
reagent syringe when attaching plunger.
Remove grey stopper on reagent syringe. Unscrew
blue cap on port No. 2 of disposable and attach
reagent syringe. While holding the Clotalyst
®
disposable vertically, inject all of reagent. Remove
the reagent syringe from the disposable. Remove
protective cover on white cap and discard. Screw the
white cap onto port No. 2.