Operating Instructions
Both battery packs provided with the Biomet
SpinalPak
®
®
Non-invasive Spine Fusion Stimulator System are partially charged
prior to being packaged. Upon receipt of the Biomet
SpinalPak
®
®
Non-invasive Spine Fusion Stimulator System, it is recommended
that you take the second battery pack, place it into the charger
cradle and charge fully. In the meantime, you may use the first
battery pack to begin your treatment immediately. Note: The first
battery pack may not provide a 24-hour treatment initially.
Step 1:
Plug the A/C adapter of the battery
pack charger into a wall outlet (Figure
4). A green light on the A/C adapter
will illuminate indicating power (Figure
5). Connect the A/C adapter and
cradle.
Figure 4
Note: At room temperature, (24°C
(75°F)), charging may take two to
three hours. In warmer or colder
temperatures, the battery may take
longer to charge.
Step 2:
Following the arrows (1, 2), insert
Figure 5
the charged battery pack into the
Biomet
SpinalPak
Non-invasive
®
®
Spine Fusion Stimulator System
(Figure 6). The LED light will blink,
indicating ready for use. Each
symbol will be indicated on the
screen and the alarm will flash
and beep if the electrodes are not
properly applied. To silence the
audio alarm, press the button
Figure 6
below the messaging screen. If the
light does not blink, this indicates the battery pack is not charged.
Charge the battery pack (see CHANGING BATTERY PACKS below).
Step 3:
Attach electrodes as per instructions on pages 3 and 8.
Helpful Tips:
Loose electrodes – Confirm that both electrodes are in complete
contact with clean, dry skin. Moisten or replace worn electrodes
if necessary.
Incomplete circuit/disconnection – Check all connection points,
confirming a snug fit.
Broken electrode lead wire – If alarm continues after confirming
connection, attach a new electrode lead wire.
4