Tab. 1: Electromagnetic Emissions
The PulsePen is suitable for use in the specified electromagnetic environment. The user of the PulsePen should
assure that it is used in an electromagnetic environment as described below.
Emissions tests
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage fluctuations /
Flicker Emissions
IEC 61000-3-3
Tab. 2: Electromagnetic Immunity
The PulsePen is suitable for use in the specified electromagnetic environment. The user of the PulsePen should
assure that it is used in an electromagnetic environment as described below.
Immunity
Test
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical Fast
Transient / Burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
Interruptions
and Voltage
variations on power
supply input lines
IEC 61000-4-11
Compliance
Group 1
Class B
Not Applicable
Not Applicable
IEC 60601-1-2
Test level
±8 KV contact
±15 KV air
±2 KV
±1 KV differential
mode
±2 KV common mode
0% U
, 0.5 cycle
T
@ 0°, 45°, 90°, 135°,
180°, 225°, 270°, 315°
0% U
, 1 cycle /
T
70% U
, 25/30
T
cycles , @ 0°
0% U
, 250/300
T
cycles
Electromagnetic Environment
The PulsePen uses RF energy only for its internal function.
Therefore, the RF emission is very low and not likely to cause any
interference in nearby electronic equipment.
The PulsePen is suitable for use in all establishments,
including domestic establishments and those directly connected to
the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Compliance
level
IEC 60601-1-2
Floors should be wood, concrete or ceramic tile.
Test level
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Not Applicable
Mains power quality should be that of a typical
commercial or hospital environment.
Not Applicable
Mains power quality should be that of a typical
commercial or hospital environment.
Not Applicable
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the PulsePen requires continued operation
during power mains interruptions, it is
recommended that the computer used with the
PulsePen be powered from an uninterruptible
power supply or a battery.