Dickson KT6P5 Snelstarthandleidingen - Pagina 6
Blader online of download pdf Snelstarthandleidingen voor {categorie_naam} Dickson KT6P5. Dickson KT6P5 9 pagina's. Remote sensing temperature recorder
Ook voor Dickson KT6P5: Aan de slag (7 pagina's)
CALIBRATION
CALIBR
CALIBR
CALIBR
ATION
ATION
ATION
The R400 Replaceable Sensor was carefully tested and calibrated before being
shipped from the factory. For greatest accuracy, we recommend replacing the
sensor every 6-12 months.
To order a Calibrated Replacement Sensor Call customer service at (630) 543-
3747 or go to www.dicksondata.com.
When your new Replaceable Sensor arrives, simply turn off the recorder, remove
and discard the old sensor, plug in the new one and power the recorder back on.
Your recorder will continue to record temperature and humidity without
interruption.
USER CALIBRATION
USER CALIBRATION
USER CALIBRATION
USER CALIBRATION
If you have an accurate standard to compare against, the Replaceable Sensor calibration
can be adjusted at one point. This will not adjust the span and is not as accurate as
replacing the Replaceable Sensor with a newly calibrated one.
1. To activate Calibration mode, turn the unit on and press and hold both the On/Off
and the Adjust
Adjust
Adjust- - - - Down
Adjust
Down button until the LED is solid green. The LED will then blink
Down
Down
amber and at which point only the pen being adjusted will show on the display.
Note:
Note:
Note:
Note: Make sure to press the Adjust-Down button first so the unit will not turn off
2. To raise the unit of measurement, press the Adjust
of measurement, press the Adjust
3. When Calibration is complete, press the On/Off
memory even after you turn the unit off or if AC power fails.
Note:
Note:
Note:
Note: After two hours, if no buttons are pressed, the unit will time out of Calibration
mode and resume normal operation. If you wish to cancel the Calibration, simply
enter Calibration mode and toggle through the steps without adjusting displayed
readings. Exit by pressing the On/Off
calibration settings.
Note:
Note:
Note:
Note: It is recommended that you use a controlled chamber when determining if
Calibration is necessary. Adjusting the pen in an open room is not recommended as
air flow can vary greatly within a very small area.
Dickson KT6P Manual V3
Adjust- - - - Down
Adjust
Adjust
Adjust- - - - Up
Adjust
Adjust
Up button.
Up
Up
On/Off button. The adjustment is stored in
On/Off
On/Off
On/Off button. You have now restored factory
On/Off
On/Off
6
Down button. To lower the unit
Down
Down
On/Off
On/Off
On/Off