Olympus BX51M Gebruiksaanwijzing - Pagina 25

Blader online of download pdf Gebruiksaanwijzing voor {categorie_naam} Olympus BX51M. Olympus BX51M 36 pagina's. System metallurgical microscope

Olympus BX51M Gebruiksaanwijzing
Problem
4. Observation Tube
a) Field of view of one eye does not
match that of the other.
5. Stage
a) Image shifts when you touch stage. Stage is not properly mounted.
b) Stage cannot travel in Y-axis direc-
tion (U-SIC4R2/SIC4L2).
c) Stage knob is too light or too heavy
to rotate (U-SVRM/U-SVLM)
Cause
Interpupillary distance is incorrect.
Incorrect diopter adjustment.
Different eyepieces are used on left and
right.
You are not accustomed to parallel opti-
cal axis.
Y-axis travel is locked.
X-axis and/or Y-axis rotation tension is
not adjusted properly.
Remedy
Adjust interpupillary distance.
Adjust diopter.
Change on eyepiece to match the
other so that both sides are the same
type.
When looking into eyepieces, do not
stare at image from the beginning but
see the overall field of view. It is some-
times recommended to turn your eyes
away from eyepieces, look far off and
look into eyepieces again.
Clamp stage.
Release lock.
Adjust it again.
BX51M
Page
15
15
14
13
13
22