5-2 Stage

@
²
³
13
Fig. 20
ƒ
...
|
Fig. 21
1
Placing the Specimen
Put the specimen in the center of the stage.
# When the specimen is placed on slide glass, turn it over so that the
cover glass faces the objective.
When using a 35 mm petri dish
}When the standard stage plate is mounted on the CK40, the 35 mm
petri dish can be mounted directly on the stage.
1. When using the CK30, put the provided 35 mm petri dish holder @ on the
stage and mount the 35 mm petri dish on the opening in the center.
2. To move the petri dish, slide the entire holder.
Using the mechanical stage
1. When using an 96-well or 24-well micro-titre plate, extend the specimen
holder to directly hold the micro-titre plate. (Fig. 21)
2. To hold any other type of plate, combine one of the following provided
holders with the mechanical stage:
· Terasaki holder ³ (AB4488) for Terasaki plate
· Petri dish holder | (AD0675) for 35 petri dish
· Slide glass holder ƒ (AB4489) for slide glass 54 petri dish
3. Turn the transverse feeding ring ... and longitudinal feeding ring † to
move the specimen to the required position (stroke: 120 mm width, 78
mm length).
2
Moving the Specimen
Turn the feeding rings on the mechanical stage or move the specimen
directly by hand to the required position.
# Be careful when changing objectives. When objectives are switched
after observing the specimen with an objective with short working
distance, the objective may interfere with the stage plate or petri
dish holder.
}When using the CK40, the IX-CP50 stage insert plate ( 50) has a wide
range of use without interference.
(Fig. 20 & 21)