Olympus CK30 Gebruiksaanwijzing - Pagina 27
Blader online of download pdf Gebruiksaanwijzing voor {categorie_naam} Olympus CK30. Olympus CK30 39 pagina's. Culture microscope
@
²
CK40-SL
@
²
CK40-RPSL
Fig. 35
³
|
Fig. 36
Centering the Light Annulus or
4
Relief Slit
# The CK40-SLP does not need to be centered.
1. Place a specimen on the stage and bring it into focus.
2. Replace the objective in the sleeve which does not have a diopter ad-
justment ring with the CT-5 centering telescope.
3. Make sure the magnification of the objective in the light path matches
that of the light annulus on the phase slider or that of the relief slit on the
RP slider.
4. While looking into the centering telescope, turn the knurled dial to focus
on the phase annulus ² of the objective corresponding to the light an-
nulus or relief slit @. (Fig. 35)
5. Insert the centering levers | into the two centering tapped holes ³
on the phase slider or RP slider. Tighten and loosen the centering
levers until the light annulus or relief slit @ is superimposed on the
phase annulus ². (Figs. 35 & 36)
6. Repeat these steps to adjust the other objectives. Keep in mind, however,
that 10X and 20X objectives share the same light annulus with the CK40-
SL and CK40-RPSL. Put an uncentered objective into the light path to
make absolutely sure the light annulus or relief slit @ is not deviating
from the phase annulus ². If there is any deviation, perform the centering
procedure again.
# Optimum performance cannot be achieved if the light annulus or
relief slit is not properly centered,
# Ghost images of the light annulus or relief slit may sometimes emerge.
If this happens, superimpose the brightest image with the phase
annulus.
# When a thick specimen is moved or replaced, the light annulus or
relief slit and the phase annulus may deviate. This can reduce im-
age contrast. If this happens, repeat Steps 1 – 5 for readjustment.
# The centering procedure may have to be repeated in order to get
the best possible contrast if a specimen slide or the bottom surface
of a culture vessel is not flat. Center the light annulus or relief slit
using objectives in the order of lower to higher magnifications.
CK30/CK40
(Figs. 35 & 36)
22