Olympus CKX41 Reparatiehandleiding - Pagina 11

Blader online of download pdf Reparatiehandleiding voor {categorie_naam} Olympus CKX41. Olympus CKX41 43 pagina's. Reflected fluorescence system
Ook voor Olympus CKX41: Overzicht (7 pagina's), Brochure & specificaties (4 pagina's), Gebruiksaanwijzing (36 pagina's)

Olympus CKX41 Reparatiehandleiding
CKX31/CKX41
<CKX41>
Part
Item
Binocular
Optical axis,
tube
exit pupil center
Pafocality
* 1) Since the binocular section of CKX41 is the same as that of CX31, basic procedure of optical axis adjustment is
performed in the same manner as CX31, which is referred to CH3-BI45 repair manual regarding the adjustment
method.
2) Nesessary jigs are partially different because UIS optical system is adopted in CKX. (Refer to the above description.)
B. INSPECTION STANDARD
Standard
CH3-BI45 repair manual
(Refer to E-2 - E-12 of
repair procedure)
CH3-BI45 repair manual
(Refer to E-13 - E-16 of
repair procedure)
B-3
Method
Set the following jigs to the microscope frame
and check the optical axis and exit pupil center.
referring the CH3-BI45 repair manual.
(1) Revolving axis
1) Standard eyepiece: KN0048 (with adapter-1)
2) Objective: PlanC10x
3) Test plate: concentric circles (KN0003) or
cross micrometer (OB-M1/100SQ)
(2) Left/right optical axis
Same as the above
(3) Absolute optical axis
1) Standard eyepiece: KN0048 (with adapter-1)
2) Standard objective: KN0041
(4) Exit pupil
1) Centering telescope (KN0029)
2) Objective: PlanC10x
Set the following jigs to the microscope frame
and check the parfocality referring the CH3-BI45
repair manual.
1) Standard eyepiece: KN0048 (with adapter-1)
2) Standard objective: KN0041
3) Focusing telescope: FT-36 or U-FT