Olympus Fluoview-1000 Gebruikershandleiding - Pagina 45
Blader online of download pdf Gebruikershandleiding voor {categorie_naam} Olympus Fluoview-1000. Olympus Fluoview-1000 50 pagina's.
![Olympus Fluoview-1000 Gebruikershandleiding](/uploads/products/server2/4283203/4283203.png)
V.M. Bloedel Hearing Research Center, Core for Communication Research
Center on Human Development and Disability, Digital Microscopy Center
Figure 15. The Stimulus Setting window for photoactivation.
1. Select a field in the sample that contains 1 or more regions of interest;
2. Collect a scan of this field with the desired acquisition settings for each channel;
3. Set the Stimulus window settings as shown in Figure 15;
a. 'Use Scanner' - choose 'Main', we don't have a SIM scanner
b. choose the ROI type and mark one or more ROIs on the Live Display window
i. each bleach ROI will have "B" appended to the ROI number
c. set the pixel dwell time for bleach scanning;
d. 'Laser' - select one or more laser lines for bleaching, and their power level
e. 'Stimulate Start Setting'
i. 'Manual' requires initiating scanning with the XY button and stopping with Scan Stop button
ii. 'Auto Stop' halts bleaching after a given number of scans or a defined time period (ms)
f. 'Activation in Series' - allows photoactivation operations in a sequence
i. 'PreActivation' - record a specified number of images before activation
ii. 'Activation Time' - Set number of scans or a defined time period (ms) for activation
iii.a set of images after bleaching can be collected
4. The time interval between scans and number of total scans (including pre-bleach and bleach scans) is
set by the Timecourse window;
May 11, 2011
Olympus Fluoview-1000 User's Guide
Select scanner
Choose entire field or ROI selections
Set dwell time
Choose laser line and power level for
bleach
Start bleach manually
Stop bleach manually or automatically after
n frames or n ms [optional]
Bleach in a series of images
Set the number of initial frames
Set number of bleach frames or bleach time
45