Olympus MX61 Montagehandleiding - Pagina 33
Blader online of download pdf Montagehandleiding voor {categorie_naam} Olympus MX61. Olympus MX61 49 pagina's. 200mm/300mm compatible semiconductor/fpd inspection microscopes
5
Command List
The following table shows the commands used with the MX61/61L.
The commands and responses of the MX61/61L employ ASCII characters. Line feed code <CR><LF> is required after any
command is added after any response. *
The detailed description on each command will be given on subsequent pages.
Command
1LOG
IN *
2)
Request
1LOG
OUT
Request
1LOG?
Query
1OB
n
Request
(n: 1 to 6)
1OB?
Query
1EAS
n
Request
(n: 0 to 3113)
1EAS?
Query
1OBA
n
Request
1CUBE?
Query
1UNIT?
Query
1FPV
ON
Request
1FPV
OFF
Request
1FPV?
Query
*
1)
Line feed command <CR><LF> consists of ASCII codes 0x0D and 0x0A.
*
2)
"
" represents a space.
*
3)
Command types
Request: Command sent by the host to the MX61/61L for requesting an operation or setting.
Query:
Command sent by the host to the MX61/61L for inquiring its status.
1)
Type *
3)
Local
remote mode switching.
Remote
local mode switching.
Check of the current local/remote mode.
Rotation of the revolving nosepiece position
to the number specified with "n" in the optical
path.
Check of the current revolving nosepiece posi-
tion (objective) number.
Setting of the aperture iris diaphragm to the po-
sition with the number specified with "n".
(n: 0 = Minimum. 1550 = Intermediate. 3113 =
Maximum.)
Check of the current aperture iris diaphragm
position.
Rotation of the revolving nosepiece position
(objective) with the number specified with "n"
in the optical axis, plus setting of the aperture
iris diaphragm to the position specified with
the AS PRESET switch.
Check of the current observation method.
Check of the presence of any unit connected to
the MX61/61L.
Setting the anti-glare setting.
Cancellation of the anti-glare setting.
Check of the current darkfield glare prevention
setting status.
Description
MX61/MX61L
Response (when successful)
1LOG
+
1LOG
+
1LOG
IN or 1LOG
OUT
1OB
+
1OB
n
(n: 1 to 6)
1EAS
+
1EAS
n
(n: 0 to 3113)
1OBA
+
1OBA
n
(n: 1 = Brightfield or optional
2 = Darkfield)
1UNIT
MX2, RVn or HS
(n: 2 or 3)
1FPV
+
1FPV
+
1FPV
ON or 1FPV
OFF
30