3. Push in the engagement adjustment lever of the exciter balancer to engage the filter in the light path.
}The angle of the engagement adjustment lever can be adjusted in the following range, only when the lever is pushed in.
Before disengaging the filter from the light path, be sure to set the engagement adjustment lever to the vertical position
(0° position).
4. Start fluorescence observation and adjust the tilting angle of the engagement lever of the balancer in the light path.
· U-EXBABG: The balance of the B/G-excitation can be adjusted. Setting the lever to the 0° position enhances the
fluorescence with the longer wavelength (closer to red), and setting it to the 45° position enhances the
fluorescence with the shorter wavelength (closer to green).
· U-EXBAUB: The balance of the U/B-excitation can be adjusted. Setting the lever to the 0° position enhances the
fluorescence with the longer wavelength (closer to blue), and setting it to the 45° position enhances the
fluorescence with the shorter wavelength (closer to green).
· U-EXBAUG: The balance of the U/G-excitation can be adjusted. Setting the lever to the 0° position enhances the
fluorescence with the longer wavelength (closer to red), and setting it to the 45° position enhances the
fluorescence with the shorter wavelength (closer to blue).
Caution
1. When the exciter balancer is not required, be sure to disengage it from the light path because flare tends to occur due to
repetitive reflections on the filter surface when the engagement adjustment lever of the exciter balancer is set in the
vertical position.
2. Be sure to set the engagement adjustment lever to the vertical position before disengaging the filter from the light path or
removing the exciter balancer.