Olympus CIX100 Gebruiksaanwijzing - Pagina 36

Blader online of download pdf Gebruiksaanwijzing voor {categorie_naam} Olympus CIX100. Olympus CIX100 46 pagina's. Cleanliness inspector system
Ook voor Olympus CIX100: Handleiding systeeminstellingen (38 pagina's), Installatiehandleiding (14 pagina's)

Olympus CIX100 Gebruiksaanwijzing
OLYMPUS CIX100 Cleanliness Inspector System
8
Best Practices for the versatile sample holders for your
CIX100 system
For your convenience, all sample holders have been designed such that a stage
limit calibration with the standard filter holder for filter membranes will also be
applicable to other sample types.
If your usual application does not use the standard filter holder, it is recommended
to do the stage calibration, which is compulsory every time the system starts, using
your preferred sample holder.

8.1 Mounting and inspection of filter membranes in the CIX100 system

A filter membrane inspection is the standard application of a CIX100 system. 6
different filter holder types for different sizes of filter membranes and backgrounds
of the membrane are available. They are all used identically.
To open the filter holder, rotate the clamp, remove any filter from the desk, place a
new filter centrally on the desk.
Then slide the filter holder into sample position 1 of the CIX100 stage insert by
placing it at the front of the stage insert and sliding it backwards.
If two filters are to be inspected in a row, place the second filter in the same way at
sample position 2 (left) of your stage insert.
To remove the holder, slide it to the front and then lift the holder.
36