The Monitor

.
When necessary, clean the exterior surfaces of the monitor with a cloth or swab dampened with warm
and a mild detergent solution. Do not allow fluids to enter the interior of the instrument.
WARNING:
disconnect it from AC mains power before cleaning.
CAUTION:
monitor in any liquid. Do not gas sterilize this monitor.
Do not touch, or rub the display panel with abrasive cleaning compounds, instruments, brushes, rough
surfaced materials or make contact with anything that can scratch the panel.
B

Probes

.
The probes are the only surfaces of this monitor that come in contact with the patient. Clean the probes after
each patient use.
Clean the monitor's probes with a commercial cleaning solution before attaching a new patient. Probes should
be cleaned until signs of wear or splitting occur. At this time a new probe is required.
If disinfecting is required, wipe the surfaces with Isopropyl alcohol or cidex and use a water rinse. When
sterilization is required, use ethylene oxide and be sure to follow hospital procedures.
Inspect the probe for wear or splitting after every disinfecting/sterilization process is complete. If wearing or
splitting of the probe is found upon visual inspection, a new probe should be used.
C

Patient cables

.
Do not autoclave the patient cables.
Operators Manual Pulse Oximeter Monitor DIGIOXI "PO-930"
Electrical shock and flammability hazard - always turn the monitor off and
Do not autoclave or pressure sterilize this monitor. Do not stack or immerse this
Digicare Biomedical Technology Inc.

SECTION 6 - Maintenance