DIGISYSTEM DSD-100 Gebruiksaanwijzing
Blader online of download pdf Gebruiksaanwijzing voor {categorie_naam} DIGISYSTEM DSD-100. DIGISYSTEM DSD-100 2 pagina's.
INSTRUCTION MANUAL FOR DRY BATH DSD-100 & DSD-200
(5)
(6)
(1) Thermometer Holder
(3) Power Switch
(5) Heater Lamp
Warnings:
* The surface is hot. Do not touch.
* Do not put any objects on the top of the instrument.
* Please keep the environment ventilated.
Steps:
1. Connect with power supply. Please make sure the power voltage is the same as the voltage shown on the
label back of the machine.
2. Insert the thermometer into the "Thermometer holder" (1).
3. Place the "Power Switch" (3) at "ON" position.
4. Adjust "Temperature Control" (4) to the required Temperature scale, and then "Heater Lamp" (5) lights up.
When the temperature reaches the setting value, "Heater Lamp" becomes off.
5. Check the Thermometer if the temperature is the same as the required temperature. If the thermometer
temperature is under the required temperature, turn "Fine Adjuster" (6) clockwise slightly; if the
thermometer temperature is over the required temperature, turn " Fine Adjuster" (6) counterclockwise
slightly. Check and adjust until the thermometer temperature is the same as the required temperature.
6. Insert the test tubes with specimen into the "Block" (2)
Maintenance and care:
Keep the machine clean all the time, so no routine maintenance is required. Cleaning can be done with a
damp cloth. Avoid the use of solvents as they may damage the product.
DIGISYSTEM LABORATORY INSTRUMENTS INC.
(4)
(3)
(2) Block
(4) Temperature Controller
(6) Fine Adjuster
(1)
(2)
(5)
(6)
(4)
www.digisystemlab.com
(3)
Block lifter
REV.3-2020