Technical passport / EDFA-1550

7.2 Instruments and Tools

1. An optical power meter;
2. A digital multimeter;
3. A standard optical fiber test jumper (FC/APC or SC/APC);
4. Some absolute alcohol and medical cotton wool.

7.3 Installation Steps

1. Before installing the equipment, please read the <User's manual> carefully and install the equipment
according to the <User's manual>. Note: For the man-made damage and other all consequence
caused by error installation that not according to the <User's manual>, we will not be responsible and
will not supply free warranty.
2. Take out the device from the box; fix it to the rack and reliably grounding. (The grounding resistance
must be < 4Ω).
3. Use the digital multimeter to check the supply voltage, make sure the supply voltage comply with the
requirements and the switch key is on the "OFF" position. Then connect the power supply.
4. Input the optical signal according to the display message. Turn the switch key to the "ON" position
and observe the front panel LED status. After the pump working status indicator turn into green, the
device is working normal. Then press the menu button on the front panel to check the working
parameters.
5. Connect the optical power meter to the optical signal output end by the standard optical fiber test
jumper, then measure the optical output power. Affirm the measured optical output power and the
displayed power are the same and have reached the nominal value. (Affirm the optical power meter is
on 1550nm wavelength test position; the optical fiber test jumper is the matched one and on the
connector surface has no pollution.) Remove the standard optical fiber test jumper and optical power
meter; connect the device to the network. So far, the device has been completely installed and
debugged.

8. Clean and maintenance method of the optical fiber active connector

In many times, we consider the decline of the optical power as the equipment faults, but actually it
may be caused by that the optical fiber connector was polluted by dust or dirt. Inspect the fiber connector,
component, or bulkhead with a fiberscope. If the connector is dirty, clean it with a cleaning technique
following these steps:
1. Turn off the device power supply and carefully pull off the optical fiber connector from the adapter.
2. Wash carefully with good quality lens wiping paper and medical absorbent alcohol cotton. If use the
medical absorbent alcohol cotton, still need to wait 1~2 minutes after wash, let the connector surface
dry in the air.
3. Cleaned optical connector should be connected to optical power meter to measure optical output
power to affirm whether it has been cleaned up.
4. When connect the cleaned optical connector back to adapter, should notice to make force appropriate
to avoid china tube in the adapter crack.
5. If the optical output power is not normal after cleaning, should pull off the adapter and clean the
other connector. If the optical power still low after cleaning, the adapter may be polluted, clean it.
(Note: Be carefully when pull off the adapter to avoid hurting inside fiber.)
6. Use compressed air or degrease alcohol cotton to wash the adapter carefully. When use compressed
air, the muzzle aims at china tube of the adapter, clean the china tube with compressed air. When use
degrease alcohol cotton, insert directions need be consistent, otherwise can't reach a good clean
effect.