Switch ON the SIREMOBIL Iso-C and release the horizontal brake (the LED in the
switch lights).
•
Move the horizontal carriage so that the opening in the horizontal carriage,
(1/Fig.14) as shown in Fig.14 is located opposite the opening in the bearing support (2/
Fig.14).
•
Insert the cable bundle (3/Fig.14) between the horizontal carriage and the bearing
support in the direction of the C-arm. When doing this, be careful not to damage or crimp
the cables!
NOTICE
•
Subsequently move the horizontal carriage carefully forward out of the bearing support
until the horizontal lift brake (1/Fig.15)with the plate (2/Fig.15) is accessible.
•
Note the following:
- Be careful not to crimp or damage the cables.
- The C-arm must remain centered on the work table. Move the work table, if necessary.
- Secure the cover with the switches to prevent it from falling.
- When removing the horizontal carriage, keep it centered to the ball bearings. Do not tilt
it.
- You may have to adjust the position of the vertical lifting column until the horizontal
carriage moves easily.
•
Switch OFF the SIREMOBIL Iso-C.
•
Disconnect the plug connections for the horizontal brake.
•
Remove the plate (2/Fig.15).
•
Remove the horizontal brake cable from the plate and lift the horizontal brake out of the
bearing support.
SIREMOBIL Iso-C
Replacing the horizontal lift brake
3
Two persons are required to remove the horizontal carriage. One person
should be at the bearing surface of the C-arm. The second person at the
horizontal carriage.
When removing the horizontal carriage, make sure that none of the
cables is crimped or damaged.
SPR2-230.076.01
Rev. 01
2
Fig. 15
Page 8 of 16
02/99
TD SD 24
1
Siemens AG
Medical Engineering