Siemens ULTRAMAT 6F Gebruiksaanwijzing - Pagina 22
Blader online of download pdf Gebruiksaanwijzing voor {categorie_naam} Siemens ULTRAMAT 6F. Siemens ULTRAMAT 6F 49 pagina's. Gas analyzer for the determination of ir-absorbing gases
Ook voor Siemens ULTRAMAT 6F: Handleiding extra uitrusting (20 pagina's), Veiligheidshandboek (24 pagina's)
4.5
Analyzers for operation in hazardous areas
4.5.1
General remarks on connecting
①
Electronics unit
②
Terminal block for electric connections
③
Cable glands
Figure 4-4
ULTRAMAT 6F, enclosure open
For all field devices, the left half of the enclosure contains the electronics unit
make the connections, you must open the enclosure of the field device. To do this, unscrew the spring bolts
the spring pressure. You can then remove the screws by hand.
4.5.2
Connecting flame arrestors
Installation of the flame arrestors
Sample gases which are occasionally explosive may only be connected to the Calomat 6F, Calomat 62F and Oxymat 6F
gas analyzers if the sample gas inlet and outlet are each provided with one flame arrestor with adapter.
You can find information on the installation of the flame arrestors in the section Connection of the flame arrestors (Page 11).
22
④
Sample gas inlet
⑤
Analyzer section
⑥
Spring bolts (example; total of 8 for the two parts of the
enclosure)
①
Compact Operating Instructions in accordance with EN 61010-1 and EN 60079-0
, the right half the analyzer section
A5E45779389002-01, 09/2018
⑤
. To
⑥
until you feel