ACD Cal 2000 Gebruiksaanwijzing - Pagina 9
Blader online of download pdf Gebruiksaanwijzing voor {categorie_naam} ACD Cal 2000. ACD Cal 2000 13 pagina's. Gas generator release vi
II.
Normal Operation: Diffusion Mode
To start the generator, press and hold the POWER switch,
located in the middle left front of the instrument, until the
display reads A. C. D., approximately three (3) seconds.
Release the switch immediately thereafter.
The instrument will sequence through several screens as
follows:
A. C. D. VI
Copyright 2000
Please Note: If you would like to change the language of
the menus, see section IV. Menu Options: Foreign
Language Option, for instructions.
The instrument will display the Serial Number and Cell
Number, followed by the manufactured date.
Serial # 0000001
Cell # 000001
Manufactured
01 Jan 2000
Calibration Date
01 Jan 2000
Page 7
V.
Maintenance
The electrochemical cells of the CAL 2000 require periodic
maintenance to provide optimum performance. It is
recomended that the electrolyte solution be replaced on an
annual basis for most gases generated by the CAL 2000.
The exception to this is the ClO2 electrolyte. Please refer to
the instructions enclosed with every chlorine dioxide cell or
replacement electrolyte pack.
If the instrument is returned with the cell on an annual basis
to maintain NIST certification, the electrolyte will be replaced
by the factory.
The following steps should be taken to replace electrolyte.
........a. Remove cell as described in section I. General
Description: Cell Installation and Removal.
........b. Unscrew solid fill plugs on reservoir and
remove liquid.
........c. Use syringe (provided with electrolyte) to add
liquid to within 1/8 inch from the bottom of fill hole.
Warning: Do not over-fill reservoir. Trapped air must be
allowed to escape through vent plug as fill screw is
replaced. Contact the factory for more information.
........d. Replace fill plugs.
Caution
Both the electrochemical cell and the alkaline batteries used
in the CAL 2000 instrument contain corrosive chemicals. In
addition, depending on the type of cell being used, the gas
being generated can be corrosive. The design of the
instrument takes this into consideration, and the 300
second purge cycle allows the corrosive gases to be swept
out of the instrument. It is important not to remove the
generating cell before this purge cycle is completed. While it
is not expected that the chemicals in the cell or alkaline
batteries will leak during normal operation, it is recom-
mended that both the generating cell and the alkaline
batteries be removed from the instrument if it is being stored
for periods longer than one week between use.
Page 16