CA.MI Clineb Pro Gebruiksaanwijzing - Pagina 7
Blader online of download pdf Gebruiksaanwijzing voor {categorie_naam} CA.MI Clineb Pro. CA.MI Clineb Pro 11 pagina's.
EN
The CLINEB PRO atomiser does not need maintenance or lubrication.
With regard to training, given the information contained in the user manual and since it is easy to understand the said device, it
doesn't appear to be necessary.
Before use always check correct functioning and safety of the device. Carry out disinfection as described in the "CLEANING
ACCESSORIES" section. Unpack the instrument and always check integrity of plastic parts and feeding cable, they might have been
damaged during previous use. Connect cable to electrical network and turn switch on.
Close the air outlet with one finger to make sure that noise produced is regular and there is no malfunctioning.
With the air outlet always closed check the correct functioning of the nebulization regulator by turning the knob form MIN to MAX.
Make sure the indicator of the pressure meter is working correctly. Verify that the atomiser is not damaged by previous use (it was
badly put away or badly knocked). A protection fuse (F 1x1.6A L 250V) reachable from exterior and it situated in the plug
protects the instrument. For use replacing, always check the type and the range indicated.
CA-MI S.r.l will provide upon request electric diagrams, components list, description, setting instructions and any other information
that can help the technical assistance staff for product repair.
Fault type
1. The device doesn't work
2. Low Nebulization
3. Low Nebulization
4. Absence of Nebulization
5. Slow Nebulization
6. Noisy Device
Fault 1 - 2 - 3 - 4 - 5 - 6
If the unit doesn't nebulizer once the above conditions have been checked, we suggest to contact your dealer or technical service
CA-MI.
The manufactured cannot be held liable for accidental or indirect damages should the device be
modified, repaired without authorization or should any of its component be damaged due to accident
or misuse. Any minimal modification / repair on the device voids the warranty and does not guarantee
the compliance with the technical requirements provided by the MDD 93/42/EEC (and subsequent
changes) and its normatives.
The device must be checked before each use in order to detect malfunctions and / or damage caused by transport and / or
storage.
During the inhalation must sit in an upright and relaxed position at a table and not in an armchair, to avoid compressing the
airways and therefore compromising the effectiveness of the treatment.
WARNING: Put the device on a flat and stable surface in order not to block the cooling vents on the sides of the device.
Extract the power supply cable and insert the plug into the mains socket.
It is recommended to unwind the entire length of the power supply cable to prevent dangerous overheating. If the power
supply cable is damage and must be replaced contact the CA-MI technical service;
Using your fingers, press the two side tabs of the nebuliser cup that hold the top of the nebuliser cup against the bottom of
the same component;
Pour the medicine prescribed by the doctor into the nebulizer;
Re-close the nebulizer, re-screwing the lid;
Connect the air tube to the air outlet nozzle of the device positioned inside the accessory compartment, making sure to keep
the cover lifted during operation;
Connect the desired accessory to the nebulizer: child mask or adult mask, mouth-piece or nosepiece;
Ensure that the supplied air filter is present;
Press the ON/OFF switch to position I to start nebulization.
To interrupt or stop the treatment press again the ON/OFF switch.
Adjust the nebulization speedy by turning the know towards MIN for longer treatments, or towards MAX for quicker
treatments.
After treatment has been completed press the ON/OFF switch to position 0 and pull out the plug from the wall socket.
Wash the nebulizers and the accessories as explained in the Cleaning section.
Place back power cord and accessories into the compartments.
MAINTENANCE
Cause
a) The plug may be misplaced in the
wall socket
b) Internal wires disconnected
c) Blocked Motor
Clogged Nebulizer Tank
Clogged Nebulizer Tank
Clogged Nebulizer Tank
Highly dense drug
Extended use
No solution with previous items
INSTRUCTION FOR USE
7
Solution
a)
Make sure the plug is properly placed in the
wall socket. Make sure the ON/OFF switch
is in position I.
b)
Contact your dealer or the CA-MI service
centre
c)
Contact your dealer or the CA-MI service
centre
Clean and disinfect the nebulizer tank as
explained in the instruction manual
If cleaning was not succesful change cruet
Check that the nebulizer contains medication;
Make sure that the nebulizer is not clogged;
Check the connection between the compressor
air outlet port and the accessories
Dilute drug in physiological liquid
Call retainer or manufacturer CA-MI
Call retainer or manufacturer CA-MI