Mode selection
TRIGGER in
OUTPUT
"Fixed Width"
OUTPUT
"Re-Trigger"
Trigger is ignored
In FIXED WIDTH mode, the NL405 will produce an Output pulse of the duration as specified on the Front Panel
regardless of further Trigger applied to the input (IN). Trigger pulses received during the time the Output (OUT)
is active are ignored.
In RE-TRIGGER mode, the Output pulse duration of NL405 is extended when a further trigger pulse is received
at the input (IN). The number of times an output can be extended by re-triggering, is unlimited.
Output "reset" and Inhibit
RANGE SWITCH
TRIGGER in
OUTPUT
Output "Delayed"
The Output pulse duration can be terminated early, in either mode, by rotating the Range control. It would be
normal, in this instance, to rotate it to the next 'click' position and then back to where it was set. The Outputs
are inhibited for 450 ms after the switch becomes still. Note that a Delayed Output pulse is NOT generated at
the end of a WIDTH that is terminated early.
Digitimer Ltd
NeuroLog™ System
Output is re-triggered here
Fig. NL405-1 : Modes - Fixed Width / Re-Trigger
Trigger ignored as within 450ms of "reset"
WIDTH
Fig. NL405-2 : Output "reset" and Inhibit
Page 2 of 5
Tel:+44 (0)1707 328347; Fax:...373153
Switch rotated here
<WIDTH
WIDTH "reset"/shortened
Output is "reset" here
NL405
450ms
No Delayed Output
Pulse, as "reset"
Copyright © 2003