7.8.

Changing C-Test Settings

Default Settings: Once you successfully communicate with the
C-Test, you will see that the unit comes programmed with a
number of default settings.
Unless necessary we do not advise changing the following
fields:
Number of Event Logs
Maximum bump time
Location customisation: This field can be used to assign a
unique name to the docking station. This will appear on the test
log file. Ex: "Gate 3 C-Test"
Auto Power-Off Time: After performing a test the unit will
power off after 120 seconds, if you wish for the docking station
to stay on longer, you can change this field.
Gas Bottle Fields: Users can program in detailed information about their calibration gas. Expiration
Date and Bottle Lot # will appear on each test record. If the expiration date is set to today's date, that
field will remain blank on the test record.
Testing Check Boxes: Changing any of these fields may disable key testing functions, please ensure
these actions are understood before changing any of the pre-set values.
Allow Calibrations – Allows a calibration if the Calibration button is pressed
Allow Bump Tests - Allows a bump test if the Bump button is pressed
Allow Hibernate – Allows the hibernation functionality described in section 2.3
Allow Configurations – Allows automatic change of configuration during test
Allow Firmware Upgrades – Allows automatic firmware update of all monitors presented
Auto cal after Bump fail – Allows a calibration to run should a monitor fail and bump test
Firmware upgrade: The C-Test allows for firmware upgrades to occur during the lifetime of the unit.
These may be product improvements to add additional features or to enhance monitor performance.
Most firmware upgrades are not required unless notified by Crowcon.
After you have made your requested changes click on the Write Device Button. If a successful write has
occurred you will see a green bar at the bottom (see above).
Issue 1 - June 2012