The MBP27T/MBP27T-2/MBP27T-3/MBP27T-4 is intended for use in the electromagnetic environment
specified below. The customer or the user of the MBP27T/MBP27T-2/MBP27T-3/MBP27T-4 should
assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
interruptions and
voltage variations on
power supply input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
General Information
IEC 60601 test
level
6 kV contact
6 kV contact
8 kV air
8 kV air
2 kV for power
Not applicable
supply lines
1 kV for input/
output lines
1 kV line(s) to line(s)
Not applicable
2 kV line(s) to earth
<5 % UT
Not applicable
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
3 A/m
3 A/m
Compliance
Electromagnetic environment –
level
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
MBP27T/MBP27T-2/MBP27T-3/
MBP27T-4 requires continued
operation during power mains
interruptions, it is recommended
that the MBP27T/MBP27T-2/
MBP27T-3/MBP27T-4 be powered
from an uninterruptible power
supply or a battery.
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital
environment.
guidance
31