Troubleshooting

Condition
Leak
Indicator changes
quickly.
10
Possible Cause
O-rings between cartridge
and manifold are not leak
tight.
Leak in gas line.
Contaminated Gas
Recommendation
Remove and check the manifold O-
rings for dirt or scratches. Inspect
the sealing surfaces on the cartridge
and manifold for scratches. Replace
if necessary.
Leak-check the overall system.
Verify that the source meets
specifications as listed under
"Limits" on page 5.
If the gas source is supplying other
filters as well, check the condition
of these filters. Otherwise check
specification and procedures for
verifying gas quality.