• Go to type file enter – all files, click to highlight then open
• Hit program – you cannot do anything whilst it is programming or you may wipe out
files. Whilst firmware is upgrading EC's screen will go blank
• Once programming complete hit close – S/N CANCELLED test complete will be
displayed on the computer
• Wait for unit to reset
• Connect – upgrade firmware No (this option will always come up)
• Once unit starts to Beep hit Next
• Check Airflow
5.2 Check Airflow
• You only need to check the airflow with the 1
• Check airflow is 10L/min (this will only come up at startup of program and each time
the program is initiated)
• The screen will also show a small window at point with "Check Airflow is between 11
and 12"
• Disregard this message and check your flow meter is correctly set to 10L/min.
REF: EC Service and Calibration Instructions V8
st
EC of the day.